EasyManua.ls Logo

laerdal Moyo - Page 18

laerdal Moyo
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Moyo is designed to monitor FHR, and provide visual and auditory alarm when
abnormal FHR is present over a prolonged period.
Moyo shows the number and status light in 3 different colors as shown below:
Simulate / tap on the sensor for 3 minutes at the rate of
e.g. 80 bpm (beats per minute) to trigger an abnormal
FHR alarm.
After you have triggered the alarm, continue to the next page.
Between
110 - 160 bpm
More than
160 bpm
Less than
110 bpm
(yellow and red)
(yellow and red)
(green)
PRACTICE
Normal FHR
Abnormally low FHR
Abnormally high FHR
Abnormal FHR alarm
To learn more about abnormal FHR alarm rules, refer to pages 36 and 48.
Yellow color turns red when abnormal FHR is present for a period of time.
1818
Alarms