EasyManua.ls Logo

LeCroy 9450 - Page 151

LeCroy 9450
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Commands 5
ACQUISITION TIME_DIV, TDIV
Command/Query
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
Response Format
The TIME_DIV command modifies the time-base setting. The
new time-base setting may be specified with suffixes NS for
nanoseconds, US for microseconds, MS for milliseconds, S for
seconds or KS for kiloseconds. An out-of-range value causes the
VAB bit (bit 2) in the STB register (Table 8, STB) to be
The oscilloscope will force random interleaved sampling (RIS) for
time-base settings in the "ON" only range, and single-shot sam-
piing for time-base settings in the "OFF" only range. For
time-base settings in the "Selectable" range, the INTERLEAVED
command allows the user to choose the required sampling mode.
RIS 9410/14 9420/24/30 9450
"ON"
2 ns/div 1 ns/dtv
~;
1 ns/div
only to to to
20 ns/div 20 ns/div
5 ns/dlv
Selectable 50 ns/div
50 ns/dlv 10 ns/dlv
"ON"/"OFF"
to to to
5 ~s/div 20 ~Ls/dlv 5 ~s/div
"OFF"
10 ~ts/div 50 ~s/dlv 10 p.s/dlv
only
to
to to
1000 s/div 5000 s/dlv 5000 s/dlv
Sequence mode acquisitions also force single-shot sampling and
therefore cannot be performed for time-base settings in the RIS
"ON" only range (see table above).
The TIME DIV? query returns the current time-base setting.
Time DIV <value>
<value> := 1 NS to 5 KS
Note: The suffix S (seconds) is optional.
Time_DIV?
Time DIV <value>
:l: 2 ns/div for Model 9430
145

Table of Contents

Other manuals for LeCroy 9450

Related product manuals