4.6 RS232C Interface
The built-in RS232C interface allows transmission of digital measurement data and
instrument conditions as well as the setting of instrument parameters.
Caution
Caution: data transmission errors
If the transmitter is operated with the RS232C interface and a fieldbus
interface at the same time, data transmission errors may occur.
The transmitter must not be operated with the RS232C interface and
Profibus at the same time.
4.6.1 Description of the Functions
The interface works in duplex mode. A nine byte string is sent continuously without
a request approx. every 20 ms.
Commands are transmitted to the transmitter in a five byte input (receive) string.
Operational parameters
Data rate
9600 Baud set value, no handshake
Byte
8 data bits
1 stop bit
Electrical connections
TxD
Pin 13
RxD
Pin 14
GND
Pin 5
(Sensor cable connector)
4.6.1.1 Output String (Transmit)
The complete output string (frame) is nine bytes (byte 0 … 8). The data string is
seven bytes (byte 1 … 7).
Format of the output string
Byte No Function Value Comment
0 Length of data string 7 (Set value)
1 Page number 5 (For ITR 90)
2 Status
Status byte
3 Error
Error byte
4 Measurement high byte 0 … 255
Calculation of pressure value
5 Measurement low byte 0 … 255
Calculation of pressure value
6 Software version 0 … 255
Software version
7 Sensor type 10 (For ITR 90)
8 Check sum 0 … 255
Synchronization