Plug-in control
42
GA 05.139/1.02 - 06/99
4.3.3 Settings pump
Menu item Description Adjustable value / option Ac-
cess
min. max. stan- Unit
value value dard
Normal Operation Threshold for normal operation
corresponding to the 35 99 95 % r/w
frequency setpoint
Max. Accel. Time Monitor time for acceleration
and overload 600 3600 1000 s r/w
Pump Type Displays the installed pump type Installed pump type - r
4.3.4 Settings temperature management system
Menu item Description Adjustable value / option Ac-
cess
min. max. stan- Unit
value value dard
TMS Setpoint Setpoint Temperature- 20 62 62 °C r/w
Management-System
4.3.5 Settings purge / vent
(Not active at MAG W 2010 CHT)
Menu item Description Adjustable value / option Ac-
cess
min. max. stan- Unit
value value dard
Purge / Vent The operator can select one Purge OFF - r/w
of the beside standing Purge ON -
options for purge and vent Controlling purge / vent
via control plug X14 -
Option P / V Displays the factory installed
option for purge and vent Installed option - r
Set Pump
Enter
Normal Operation
95 %
Enter
Max. Accel. time
800 s
Enter
Pump Type
MAG 1600/2000 CT
Enter
Set TMS
Enter
TMS Setpoint
60 °C
Enter
r (read) = value can only be read
r /w (read/write) = value can be read and written
Via X14
Purge ON
Set Purge/Vent
Enter
Purge/Vent
Purge OFF
Enter
Option P/V
Option 1
Enter