1.3.2 Technical Data PHOENIX L300i DRY
Physical data
To get down to the minimum detected leak rate range some conditions must be
fulfilled:
PHOENIX L300i has to run at least 20 minutes
Ambient conditions must be stable (temperature, no vibration/accelerations.)
The part under test has been evacuated long enough without using the zero func-
tion (background is no longer decreasing)
ZERO must be active
Max. inlet pressure 15 mbar
Minimum detectable Helium leak rates
- in vacuum mode
< 3·10
-11
mbar l / s
- in sniffer mode
< 1·10
-7
mbar l / s
Minimum detectable Hydrogen leak rates
- in vacuum mode
< 1·10
-8
mbar l / s
- in sniffer mode
< 1·10
-7
mbar l / s
Maximum Helium leak rate which can be displayed 0.1 mbar l / s
Measurement range 11 decades
Time constant of the leak rate signal (blanked off,
63% of the final value)
<1 s
Pumping speed (Helium) at the inlet
GROSS mode 0.02 l / s
PRECISION mode 0,4 l / s
FINE mode > 2.5 l / s
Detectable masses 2, 3 and 4
Mass spectrometer 180° magnetic sector
field
Ion source 2 filaments;
Iridium/Yttrium-oxide
Inlet port DN 25 KF
Run-up time (after starting) < 2 min