Operation
GA10218_002_C0 - 10/2016 - © Leybold GmbH 75
3.10.3 Mass
Main menu > Settings > Mass
The requested mass of the measured gas can be selected. The PHOENIX L300i
must be in standby mode for changing to another mass.
After changing the mass a calibration for the selected mass should be done. See
chapter
3.13.
3.10.4 Miscellaneous (Language, Calibration
request, Service interval...)
Main menu > Settings > Miscellaneous
The actual date and time, the preferred language and the mains frequency can be
set in this submenu.
Time & Date
Main menu > Settings > Miscellaneous > Time & Date
Time and date can be changed on two subsequent pages. Please refer to
Numerical entries for the description of the entry.
Softkey 2: H
2
(2 amu)
Hydrogen with the mass of 2 amu will be measured.
Softkey 3:
3
He (3 amu)
Isotope of helium with the mass of 3 amu will be measured.
Softkey 7:
4
He (4 amu)
Helium with the mass of 4 amu will be measured. Default setting
Softkey 2: Time & Date
Setting of time and date
Softkey 3: Language
Selection of the language
Softkey 4: Calibration request
Setting if the PHOENIX L300i should remind for a calibration
Softkey 5 Service interval forepump
Setting service time interval forepump
Softkey 7: Service interval exhaust oil filter
Setting service time interval exhaust oil filter
Softkey 8: Service message exhaust oil filter
Setting service message exhaust oil filter