18 GA01208_002_C0 - 11/2016 - © Leybold
Installation
3 Installation
Instructions for the operating personnel
Before any kind of assembly work the personnel must be informed about
to potential hazards. Observe Safety Informations 0.1 to 06.
3.1 Placement
Explosion risk
The standard pump is not suited for installation in the explosion hazard
areas. When planning such an application please contact us first.
Observe Safety Informations 0.5.
The TRIVAC-B-DOT pump can be set up on a flat, horizontal surface. Rubber
feet under the coupling housing ensure that the pump can not slip.
The location of the TRIVAC-B-DOT should be such that all controls are easily
accessible.
Risk of toppling
Max. tilt for the pump is 10° from the vertical.
If you wish firmly install the pump in place, insert bolts through bore holes in
the rubber feet.
The rubber feet act as vibration absorbers. They must therefore not be
compressed by screws.
The site chosen should allow adequate air circulation to cool the pump
(keep front and rear unobstructed).
The oil-level glass must be visible.
Note the ambient conditions during operation
The ambient temperature should not exceed +40 °C (104 °F) and not
drop below +12 °C (55 °F) (see Section 4.3.3).
For installation sites over 1000 m above sea level and/or a relative atmos-
pheric humidity of over 90%, you must discuss this with Technical Sales.
Maximum heat dissipation corresponds approximately to the installed
motor power.
When failing, capacitors may heat up excessively and generate smoke.
Therefore the pump must be placed with a sufficient clearance from
inflammable or combustible material. There exists the risk of fire!
DANGER
CAUTION
NOTICE
NOTICE
NOTICE