EasyManua.ls Logo

LEYBOLD TRIVAC D 16 B-DOT - 5.4 Oil Change

LEYBOLD TRIVAC D 16 B-DOT
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32 GA01208_002_C0 - 11/2016 - © Leybold
5.4 Oil Change
Pump and operating agents may be contaminated
Hazardous substances may escape from the pump and the oil. If there
is the danger that the operating agent may present a hazard in any way
due to decomposition of the oil, or because of the media which have
been pumped, you must determine the kind of hazard and ensure that all
necessary safety precautions are taken.use gloves, face protection or a
respirator, for example.
Observe Safety Informations 0.3 und 0.4.
Change the oil while the pump is cold so as to avoid releasing adsorbed
gases. If releasing adsorbed gases is no problem, change the oil after the
pump has been switched off and while the pump is still warm.
Risk of suffering burns
With the pump warm from operation, both pump and oil can get so hot
that there is the risk of suffering burns. If required, wear gloves.
Check and top up oil only after having shut down the pump first.
When disposing of waste oil, observe the applicable environment
protection regulations!
When changing the oil use the same type of oil which was previously in
the pump. If you want to change the type of oil entirely please consult us
first.
Maintenance
WARNING