Operation
74
300554859_002_C0 - 11/2016 - © Leybold
Fig. 4.18 Data list
Data List
Through the data list, all recorded data points can be viewed. Double-clicking
on a line with the data records then displays the corresponding point within
the graph. Moreover, all occurred events are listed in the right-hand window.
LoggingCycle
See Section Webserver DataViewer LoggingCycle
Memory Capacity
For an estimate as to how long the device will be capable of recording data,
refer to the following formula:
The log file has a size of 3264512 bytes 3264512/32 = 102016 entries in
the log file.
102016 entries x 1 second (LoggingCycle) = 102016 seconds.
From this there results a recording duration of approximately 1700 minutes
which is roughly equal to 28 hours.
The recording duration will vary depending on the selected LoggingCycle.
Device List
When wanting to remove any added devices from the device list, remove
these entries from the file “device.txt”. This file can be found in the
DataViewer Folder.