Connection
2.8 Installing the
TURBOTRONIK
For installation in a rack use the mounting frame 19“, 3
HU!
The heat dissipation of the TURBOTRONIK must not be
obstructed. Insure a sufficient ventilation - the ambient
temperature during operation must not exceed 45°C
(113°F). Make sure that there is sufficient heat dissipati-
on for the two side profiles. As a general rule, it is not
possible to install two TURBOTRONIK NT 20 units in
one 19“ rack.
If - after installation - the rear of the TURBOTRONIK
is no longer accessible, switch on the power switch (3/1)
and the DIP switches before mounting and if nec. set the
starting delay; refer to section 3.1.
Warning
If the TURBOVAC is built into a rack the
mains plug is not within easy reach.
Therefore install a separation between
the TURBOTRONIK and the mains
when you build it into a rack.
Warning
Do not operate the TURBOTRONIK with
the standard mains lead in chemically
aggressive surroundings. If you operate
the TURBOTRONIK in chemically
aggressive surroundings replace the
mains lead by a resistant one.
24
GA 05.208/12 - 05/2003