CALIBRATION cont.
If FS = 6
Please follow the procedure below:
● The display will show [dP 1], to indicate how many decimal places you want; press
UNIT to change, and press HOLD to confirm (ex. DP 2 = ####.##)
● The display will show [d 2], to indicate the division; press UNIT to change, and
press HOLD to confirm
● The display will show [CAP ], to indicate the max. capacity; press HOLD to
display the current max. capacity; press ZERO to left shift and UNIT to change the
capacity; Then press HOLD to confirm
● The display will show [LoAd ], to indicate the calibration loading weight; press
HOLD to display the current calibration loading weight; press ZERO to left shift and
UNIT to change the calibration loading weight; Then press HOLD to confirm
● The display will show [CALon] please follow the steps for FS values of 0-6 to
complete calibration
For FS values of 0 - 6
Please follow the procedure below:
● The display will show [CALon], press HOLD to enter into weight calibraion, or
press ON/OFF to exit calibration
● The display will show [noLoAd] for zero calibration; make sure there is no load
on scale, then press HOLD to confirm and zero calibrate
● The display will show [AdLoAd] for loading calibration; place calibration weight as
defined by the tables above on the scale
● Once the weighing platform is stable, press HOLD to confirm and calibrate
● The display will show [CALEnd] when the calibration is complete, press HOLD to
save the weight calibration and press ON/OFF to exit and go back to weighing mode