Operating Instructions MA 42
Operating_Instructions_8100460-6_MA42_e_15a 17 8100460-6 1/15
If you are using bone conductor with contra lateral masking transducer like the
BKH 10, connect the contra phone to the left insert phone socket (14, Fig. 1).
If masking is switched on, the masking noise will be always routed to this
output.
For hygienic reasons it is important to clean the bone conductor (see chapter 8
Disinfection).
5.1.3 Sound Field Testing
Set the receiver selector to Speaker. Perform the test the same way as for air
conduction.
Warble tones should be used in the sound field since pure tones may not
provide reproducible results in the typical test room.
Perform the test the same way as described in the air conduction section above.