2
Camel Quick Start Guide
Mangar Health
Tel: +44(0)1544 267674
Email: sales@mangarhealth.com
Web: www.mangarhealth.com
Cleaning and Disinfecng
The Camel should be fully inated and pre-cleaned by washing with detergent and water,
thoroughly rinsed and then disinfected using a 1:10 soluon of household liquid bleach and
water (1 part bleach to 10 parts cold water). Leave the soluon on the Camel for approximately
10 minutes before thoroughly rinsing o with clean cold water. Allow to air dry before storage.
Do not dry using hot air.
BATTERY NOTE: Charge for 12 hours before rst use. Before each use, check the baery charge
state by pressing the ‘Baery State Indicator’ buon (please read full Airo 24 User Instrucons).
WARNING: While in use, never allow a secon to become parally deated while any secons
above it are inated. Secons must always be fully inated from the boom up, in sequence, in
order to maintain stability.
Press Baery
State Indicator
Buon
RED
Recharge
Immediately
2 GREEN
Recharge
soon
3 GREEN
OK to Use
4 GREEN
OK to Use
1 GREEN
Recharge
before next use