45
Troubleshooting:
• If the battery gauge starts ashing during the treatment, the
charge level of the batteries is getting low. Stop the treatment
session and recharge the device.
• If the signal feels week or intermittent and / or device says “Check Leads and Pads”
- Check to make sure all pads have a good connection to the skin.
- It may be time to change to a new set of electrodes.
(see electrode care)
• If the unit is powered on and the LCD screen does not turn on, check that
the charger is unplugged from the wall and the back of the device.
• If this had been done and there is no change or if the unit does not seem
to be holding its 10 hour charge time between treatments; please contact
Marc Pro. The battery may need to be replaced.
Cleaning Your Unit:
To clean your Marc Pro Plus, use a cloth and
an alcohol-based cleaning product, which
does not contain solvents.
Technical Data:
• Voltage 0-35V • Current 0-35mA
• Frequency 1-70Hz • Bi-phasic waveform
• Pulse Duration 5ms; 1ms at 50% of max amplitude
* tests conducted under 1K load
NOTE OPERATING INSTRUCTIONS
Storage & Transport Conditions:
The Marc Pro Plus contains rechargeable
batteries and so the storage conditions
must not exceed the following gures:
Storage temperature:
from -4° F to 113° F, -20° C to 45° C
Max. relative humidity: 75%
Atmospheric pressure:
from 700 hPa to 1060 hPa
Disposal: For environmental protection the
device, the battery and its accessories
have to be disposed of properly.
Type BF Applied Part
Visit www.marcpro.com
for more info, tips, tricks and best practices
!