1312
0.7 0 100 120 No CH1 & CH2 Positive
1 0 125 120 No CH1 & CH2 Positive
1 0 125 120 No CH1 & CH2 Positive
0.7 0 100 120 No CH1 & CH2 Positive
1 0 75 120 No CH1 & CH2 Positive
0.7 0 50 120 No CH1 & CH2 Positive
0.7 0 40 120 No CH1 & CH2 Positive
0.7 0 30 120 No CH1 & CH2 Positive
0.8 0 20 120 No CH1 & CH2 Positive
1 0 15 120 No CH1 & CH2 Positive
1.4 0 10 120 No CH1 & CH2 Positive
1.5 0 8 120 No CH1 & CH2 Positive
P11 Modulation 2-100Hz 260µs-150µs Continue
Rate & width
over 6 seconds
P12 Modulation 80< - >7Hz 260µs Continue
Rate & width
over 6 seconds
Program The pre-set parameters of the EMS 13-24
Details programs are as given below:
Mode SYN/ Rate Width Ramp ON Time Off Time Timer
ALT (Hz) (µs) (Sec) (Sec) (Sec) (Min)
P13
ACL repair/joint SYN 35 300 3 8 24 20
protection back muscle
P14
Spasm small SYN 80 300 3 10 5 20
muscle
P15
Spasm SYN 80 250 2 8 4 20
Postoperative
P16
Arthroscopy SYN 25 200 2 6 30 15
P17
Disuse atrophy SYN 35 300 2 5 15 30
P18
Shoulder SYN 50 300 5 15 50 15
Subluxation
P19
Range of motion SYN 40 250 3 6 21 30
muscle re-education of hips
P20
Muscle training SYN 50 250 2 10 10 20
P21
Muscle training SYN 50 250 2 14 14 20
P22
Muscle training SYN 35 400 2 10 10 20
P23
Muscle training ALT 50 250 2 10 10 20
P24
Muscle training ALT 50 250 2 10 14 20
23 Timer
Adjustable, from 5 to 60 minutes minutes
and continue(C), 5 minutes each step
24 Patient This unit can store 60 sets of operation
Compliance Meter records. Total recorded time is 999 hours.
25 Low Battery A low battery indicator will show up when
Indicator the battery is low.
26 Operating Temperature:0°~40°C
Condition Relative Humidity: 30%~75%
Atmosphere Pressure : 700Hpa~1060Hpa
27 Remark There may be up to a +/-10% tolerance of
all parameters and +/-20% tolerance of
output amplitude & voltage.