Electromagnetic Compatibility
In order to regulate the requirements for EMC (electromagnetic compatibility) with the aim of preventing
unsafe product situations, the EMC EN60601-1-2 standard has been implemented. This standard defines
the levels of immunity to electromagnetic interference as well as maximum levels of electromagnetic
emissions for medical devices. Medical devices manufactured by Mascot have been tested and conform
with the requirements of IEC/EN 60601-1-2, 3rd & 4th edition, nevertheless, special precautions may
need to be observed:
The Mascot products are suitable for use in Domestic, Residential, Office and Hospital environments,
except in special locations where EM Disturbances are known to be high, such as near High Frequency
Surgical Equipment or Magnetic Resonance Imaging systems.
When used according to its specification the User can expect the product to fulfil its essential perfor-
mance, being powering Medical Electrical Devices or charging batteries for Medical Electrical Devices.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be
avoided because it may result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers and cables other than provided by the
manufacturer could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
WARNING: Portable RF communications equipment should be used no closer than 30 cm (12
inches) to any part of the power supply or battery charger, including cables. Otherwise, degra-
dation of the performance of this equipment could result.
GUIDANCE AND MANUFACTURER’S DECLARATION
The Masot products are intended for use in the electromagnetic environment specified below. The custo-
mer or user should ensure that it is used in such an environment.
Test / Standard Compliance level Guidance
Emission:
RF emissions,
CISPR 11
Group 1, Class B
Suitable for use in all establishments, inclu-
ding domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buil-
dings used for domestic purposes.
RF emissions not likely to cause any interfe-
rence in nearby electronic equipment.
However, a separation distance of 30 cm
shall be maintained.
Harmonic emissions,
IEC 61000-3-2
-
Voltage fluctuations /
flicker emissions,
IEC 61000-3-3
-

Related product manuals