8. The screen view can be used to conrm the correct insertion depth
of the E.T. tube.
*It is important not to advance the blade too deeply in order to
maintain maximum space to facilitate the E.T. tube placement.
**Reference to these numbers can be useful during training to avoid
inserting the blade too far.
Cleaning and Disinfection
After unpacking, decontaminate the device and battery before use.
The device and battery should be cleaned and disinfected separately
after each patient use. Cleaning and disinfection should be carried out
as per this instructions for use document and in accordance with local
regulations. Follow hospital policy and protocol when handling and
cleaning soiled items.
The instructions provided have been validated by Aircraft Medical as
being capable of preparing the device for re-use.
The following methods have been approved:
Note: Blades are not intended to be cleaned or disinfected.
Product Cleaning
Intermediate Level
Disinfection
McGRATH™
MAC with
McGRATH™
3.6 V Battery
removed
Quaternary
ammonia
+ isopropyl
alcohol (IPA)
(e.g. Super
sani-cloth or
equivalent)
1. Quaternary ammonia
+ isopropyl alcohol
(e.g. Super sani-cloth or
equivalent)
Product Cleaning Low-Level Disinfection
McGRATH™
3.6 V Battery
70% IPA wipe 70% IPA wipe
Dispose of wipes in accordance with local regulations and hospital
policy.