Other Settings and Functions:
- Away Mode You may turn the dispenser OFF by pressing the hour button and holding for approximately
4 seconds. The display will read ‘OFF’
and the medication will stop dispensing.
To return to normal
operation, press the hour until the PRN time displays.
Alarm Test Press and release this button on the upper left to test the dispensing rotation. Slide the white
T-BAR up reset the PRN time.
Trouble Shooting
No PRN time is being
displayed, but the colon is
blinking
The Device is not getting A/C power. Try another A/C wall outlet that you
know is live. Insure both ends are connected.
The PRN Time does not
reset when opening the
medication door.
Instruction # 6 was not followed when replacing the blue lid. The
medication door pushes up the white T-BAR (directly below the medication
door), which triggers a micro switch and resets the PRN time. If the
medication door is on the wrong side of the white T-BAR, the medication
door can’t engage the micro switch to reset the PRN time.
The medication tray does
not rotate when the PRN
time reaches 0:00
The battery is not plugged in. See step # 1. To test the battery connection,
disconnect the A/C power, and ONLY the colon should be blinking.
The new medication slot is
empty
The medication was not loaded properly, Refer to step #5.
Display shows ‘OFF’ Depress the Hour button by itself for several seconds to return to show the
PRN time.
Blue Lid won’t go back on See # 6 above
The Battery and A/C are
connected but the device
won’t rotate to the next
medication when the PRN
time reaches 0:00
Most likely, the medication tray was previously forced to the next dose
disabling the device. Please call our customer support at 310.328.7557 (9-
4:30 M-F Pacific Time) to rectify this situation.
Please read and understand the following:
Note that many PRN regimens involve the dispensing of narcotics. The MedReady
1700PRN is designed as a medication aid only. MedReady is NOT a medication Safe or
Vault. With determination, a patient will be able to break into the device and access more
than intended. If there is any doubt that a patient will have an uncontrollable urge to
access more medication than prescribed, please do NOT use this device. Under normal
use, only one dose is available at any one time.