Device Type | Continuous Glucose Monitoring (CGM) System |
---|---|
Manufacturer | Medtronic |
Sensor Life | 7 days |
Connectivity | Bluetooth |
Warm-up Period | 2 hours |
Mobile App Compatibility | iOS and Android |
Alerts | Customizable high and low glucose alerts |
Water Resistance | Water-resistant (not fully waterproof) |
Calibration | Requires fingerstick calibration 2 times in the first 24 hours, then once daily |
Data Sharing | Yes (with caregivers and healthcare providers) |
A complete transmitter kit includes the transmitter, two testers, and a charger.
Transmitter for single or multiple-patient use in Medtronic CGM systems.
Avoid strong magnetic fields (MRI, diathermy).
Refer to sensor guide. Choking hazard for small parts.
Use specified sensors/testers. Avoid liquids and twisting.
Avoid exposure of transmitter to MRI, diathermy, or other strong magnetic field devices.
Device complies with FCC Part 15 rules, subject to two conditions for operation.
Medtronic declares product conforms to essential requirements of Directive 1999/5/EC.
Special precautions regarding Electromagnetic Compatibility (EMC) for body worn devices.
Contact your local representative for assistance.
Step-by-step instructions for installing a battery in the charger.
Detailed steps and indicator light meanings for charging the transmitter.
Refer to system guide for pairing transmitter to mobile device.
Refer to serter guide for sensor insertion instructions.
Before proceeding, have system user guide available. Steps to connect transmitter to sensor.
Before proceeding, have system user guide available. Steps to disconnect transmitter.
Transmitter and sensor form a waterproof seal up to 2.4m (8ft) for 30 minutes.
Tester checks transmitter function and is required for cleaning. Usable for one year.
Example of connector pins. Check for damage or corrosion.
Before proceeding, have system user guide available. Steps to connect tester.
Transmitter for personal or healthcare facility use. Single-patient requires cleaning after each use.
Always clean transmitter after each use. Materials needed.
Transmitter usable up to 122 cycles or one year.
In healthcare facilities, clean and disinfect after each use. Follow standard precautions.
Procedure to remove adhesive residue using medical adhesive remover.
Steps for removing adhesive residue.
Procedure for general cleaning of the charger. Do not immerse in water.
Store devices in a clean, dry location. Recharge transmitter at least every 60 days.
Dispose of the transmitter according to local regulations for battery disposal.
Technical specifications for the transmitter and charger.
Transmitter and app connect via BLE. Transmitter sends glucose data and alerts.
Transmitter accepts RF communications only from recognized devices. Sensitive data is encrypted.
Transmitter intended for use in specified electromagnetic environment.
Transmitter intended for use in specified electromagnetic environment.
More details on immunity tests and environmental guidance.
Maintain minimum distance between RF equipment and transmitter based on output power.