MiniMed
™
780G system changing the Silhouette
®
infusion set quick reference guide
Start here
1 2 3 4
Fill reservoir & connect to the infusion set tubing
5 6
Wash your hands.
Press . Select Options.
Select Reservoir & Set. Select New Reservoir. Remove the infusion set you have been using by loosening the adhesive
and pulling away from body.
Remove the used reservoir from the pump. Select Rewind.
Follow the next steps to ll reservoir with insulin and connect to the infusion set tubing.
1 2
Plunger
rod
Reservoir
Transfer
guard
Plunger
Remove from package. Make sure insulin
vial is at room temperature to reduce the
risk of air bubbles.
Pull plunger down to
the amount that you
plan to ll with insulin.
3
Insulin
vial
Continued on next page
1
4
Wipe vial with alcohol.
Place vial on table.
Firmly press the blue
transfer guard onto vial.
Push and hold plunger down.