Preparation for use | EN
11
5. Preparation for use
5.1. Transport and storage
Thedeviceisshippedintwopolystyreneshellsxedtogether.Thisoriginalpackagingmustbeusedforanyonwardandreturntransport.Thedevicemustalways
be transported and stored in an upright position. The device must be transported to the installation site in a completely packed state. After unpacking the device,
check for completeness and possible transport damage.
70 °C
0 °C
Transport and storage temperature
80 %
0 %
Transport and storage humidity limitation
5.2. Installation requirements
40 °C
10 °C
Operating temperature: 10°C to 40 °C
70 %
0 %
Humidity limit max. 70 %
≤3000m
Max. 3,000 m above sea level
> The following space requirement must be provided for the installation: H x W x D = 350 mm x 215 mm x 120 mm
> To avoid vibrations, use only the three holes on module 1 to attach to load-bearing parts of the treatment unit! (Do not remove vibration inserts!)
> The device must be able to swing freely and must not have any contact with other parts of the treatment unit (cables, hoses, covers, etc.).
Caution:
Acontactprotectionplatemustbettedbehindmodule1ifthesubsurface
is electrically conductive!