3 Installation
IC equipment: Inline-Ultrafiltration
■■■■■■■■
7
Remove the 3 green stoppers.
Turn the ultrafiltration cell around and place it on a table. The
screws face upwards.
2
Removing the screws
Loosen the screws with the hex key.
Remove the screws with the washers and put them aside.
3
Disassembling the ultrafiltration cell
Remove the upper chamber of the ultrafiltration cell.
Remove the sealing ring.
4
Cleaning the ultrafiltration cell
CAUTION
Damage to the ultrafiltration cell
Organic solvents (e.g. acetone) corrode and damage the ultrafiltra-
tion cell material (PMMA).
Use only ultrapure water or a water-ethanol mixture (70:30) for
cleaning the ultrafiltration cell.