4.20 Connecting the separation column
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
930 Compact IC Flex ChS/PP/Deg (2.930.1360)
Start manual control in MagIC Net and select the high-pressure
pump: Manual Manual control Pump
Flow: in accordance with column leaflet
On
Rinse the guard column with eluent for approx. 5 minutes.
Stop the high-pressure pump in the manual control in MagIC Net
again: Off.
4.20 Connecting the separation column
CAUTION
New separation columns are filled with a solution and sealed with stop-
pers on both sides. Before inserting the column, ensure that this solu-
tion can be mixed with the eluent being used (follow the information
provided by the manufacturer).
NOTE
Connect the separation column only after the initial start-up of the
instrument. Until that point, insert a coupling (6.2744.040) instead of
the guard column and separation column.