3040DCT EM-9538-B Rev.5
MG CO., LTD. www.mgco.jp
5-2-55 Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
4
Press SHIFT button to show the setting and SCALE/↑ button to shift
the display into the setting standby mode.
• The forth digit starts blinking, to which you can apply changes.
5
Press DISP/↓ or SCALE/↑, and UP buttons to set to ‘1.000’.
• Press D
ISP/↓ or SCALE/↑ button to go to the next digit, and UP button to
change the blinking value.
NOTE
• ‘1.000’ is a display example. Set any value within the setting range.
• ‘ERR’ starts blinking back in Measuring Mode when the set value is within invalid range, or is same as or less than the
input scaling value A. Return the setting within the valid range.
6
Press SHIFT button to apply the new setting.
• And the setting item is indicated.
7
■ TO GO ON TO SET THE DISPLAY SCALING VALUE A,
Skip to Step 3 in “3.4 STEP 4. DISPLAY SCALING VALUE A”.
■ TO QUIT,
Hold down D
ISP/↓ or SCALE/↑ button for 1 second or more to return to Measuring Mode.
Blinking
Blinking