12.2 INITIALIZING SETTING VALUES
To restart setting from the default state, initialization can be used. Refer to attached 13.3 PARAMETER LIST for the default
values.
IMPORTANT
• Currently set parameters will be lost after an initialization. It is recommended to record the parameters before initializa-
tion.
• Even if the unit is shipped with the specied parameters with the option code ‘/SET’, such parameters will be lost after
an initialization. Be careful that the initialization does not recover the ex-factory settings.
12.2.1 OPERATING PROCEDURE
1
Hold down DISP/↑ button for 3 seconds or more to move on to
Display Setting Mode.
• The setting item ‘AVG’ (moving average sampling No.) is indicated.
2
Press DISP/↓ or SCALE/↑ button to go to the initialization.
• The setting item ‘RST’ (initialization) is indicated.
3
Press SHIFT or UP button to select ‘ON’.
1,2,
5
2,5
3,4