4.1.2 OPERATING PROCEDURE
NOTE
The left gure shows a display example (default value of
input code ‘SA’). The display depends on the specications
and settings.
1
Conrm the wiring, and turn on the power.
• All the indications turn on for approximately 3 seconds and then the
display moves on to Measuring Mode.
NOTE
Indication ‘S.ERR may blink, which shows the input out of the permis-
sible range and does not show the unit failure.
2
Hold down SC button for 3 seconds or more to move on to Scaling
Setting Mode.
• The display scaling value A is indicated.
• The rst decimal point starts blinking.
NOTE
Skip to Step 6 if the default value is acceptable.
3
Press SF button to shift the display into the setting standby mode.
• The forth digit starts blinking, to which you can apply changes.
5
2,5
3,4
4
Immediately after power on (all indicators on)
Measuring Mode
*1 Display depends on the settings and input.
Blinking
Blinking