MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT
REPAIR SERVICE INSTRUCTIONS — CALIBRATION
2266-20 Laser TEMP-GUN™ Thermometer
Environmental
Condition
Perform all calibration at an ambient temperature of 23°C ± 2°C / 73.4°F ± 3.6°F and relative
humidity of 80% - allow the TEMP-GUN™ thermometer to sit at this temperature for a
minimum of 30 minutes before proceeding.
REPAIR SERVICE
INSTRUCTIONS
Calibration
MILWAUKEE ELECTRIC TOOL CORPORATION
13135 WEST LISBON ROAD BROOKFIELD, WISCONSIN 53005-2550 USA
58-92-2266d1 04/2010
Question and Answer IconNeed help?

Do you have a question about the Milwaukee 2266-20 Laser TEMP-GUN and is the answer not in the manual?

Milwaukee 2266-20 Laser TEMP-GUN Specifications

General IconGeneral
TypeInfrared Thermometer
DisplayLCD
Battery Type9V
Laser TypeSingle Laser
Resolution0.1 °C/°F
Response Time500 ms
Auto Shut-offYes
Battery LifeApprox. 12 hours
Operating Temp.32°F to 104°F (0°C to 40°C)
EmissivityFixed at 0.95

Summary

Introduction

Precautions and Safety Information

Information on hazards, warnings, and cautions for safe operation and handling of the device.

The Symbols

Explanation of symbols used in the manual and on the device for clear communication.

Safety Precautions

Specific Precautions

Detailed safety instructions regarding laser light, radiation, and operational conditions to prevent injury.

Specifications

General Specifications

Technical details including laser type, power, wavelength, and temperature measurement ranges.

Basic Maintenance

Loading/Changing the Batteries

Step-by-step instructions for inserting and replacing the device's batteries.

Maintaining Tool

Advice on regular maintenance and when to contact a service facility for repairs.

Cleaning the Laser Windows

Instructions for cleaning the device's laser windows for optimal performance.

Cleaning the Tool

Guidelines for cleaning the tool's exterior and avoiding harmful cleaning agents.

Scope and Procedures

Equipment Use

Lists recommended reference thermometers and IR black bodies for calibration.

Testing Condition and Interference

Specifies ambient conditions, acclimatization, and factors affecting calibration accuracy.

Field Calibration Procedure

Outlines the steps for performing IR software calibration checks and KTC calibration.