Maintenance Manual > Test Items > Test Function > [HEATER TEST] TEMPERATURE
Model CJV30/TPC Issued 2008.08.04 Revised F/W ver. 1.00 Remark
1.0
5.1.11 [HEATER TEST] TEMPERATURE
Function
Temperature tests of the media heater are executed.
Operation Procedures
Step Item Description Remarks
1 Temperature setting Sets temperature of Pre, Print, and After Heater to control the heater.
Set value (Celsius): OFF, 20 – 50 C (unit: 1C)
Set value (Fahrenheit): OFF, 68 – 122 F
(as converted from Celsius, unit is not F)
2 Temperature display [ENTER]: Returns to temperature setting.
Temperature is displayed with a unit selected in the [UNIT SETUP] of the [MACHINE SETUP] function.
A/D conversion value is also displayed.