Using the QC Programs
8-4
shows.
Figure 8-5 Entering “Exp. Date”
Entering the expected results (mean) and limits (range)
ENTER the expected results (mean) and limits (range) respectively into the “Mean and
Range” boxes of the parameters to be included in the QC analysis, as Figure 8-6 shows.
Figure 8-6 Entering “Mean” of RBC
z Refer to the instructions of use of the control for information on the lot
number, expiration date, open-vial stability days, expected results and limits.
z The entered expiration date should be either the expiration date printed on
the labeling or the open-vial expiration date. It is earlier.