Using the Calibration Programs
9-13
If all the parameter values obtained are numeric, a message box will pop up to confirm the
validity of the results, as Figure 9-14 shows. CLICK Yes” to save the results to the “Fresh
Blood” screen; CLICKNo” to abort the result.
Figure 9-14 A message box to confirm the validity
Repeat the above steps to run the sample 3 to 5 times (5 is recommended) and the analyzer
will automatically calculate the CV and calibration factor, as shows. Note that the CVs should
be within the ranges specified in Table 9-1.
Figure 9-15 Calibration with fresh blood
The calculated calibration factors should be within the 75 to 125. Any calculated value
that falls out of the calibration range will be flagged with a “*” at the upper right corner. Other
values will not be displayed. In case of an empty calibration factor, try to find out the reason
and if necessary, contact Mindray customer service department or your local distributor.