Using the Calibration Programs
9-16
Figure 9-18 The dialog box
If the valid results are less than three (the CV and new factors are not available yet).
If you press [F3], a dialog box will pop up to warn you about the data loss, as Figure 9-19
shows.
Figure 9-19 The dialog box
To switch the modes, CLICKYes” and the saved data will be cleared; Otherwise, CLICK
No” to exit.
Printing new calibration factors
Press [PRINT] to print out the new calibration factors.
Exiting the ”Fresh Blood” screen
Press [MENU] to exit to system menu.