Using the Calibration Programs
9-19
Figure 9-23 A message box to warn invalid input
Seeing the box, CLICKYes” and re-enter the factors. If the changed factors are all within the
calibration range, a dialog box will pop up to remind you to save the new factors, as Figure
9-24 shows. CLICK Yes” to save the new calibration factors
Figure 9-24 A message box to confirm the new calibration factors
At the “Count” screen, run the calibrator or a normal-level control at least 3 consecutive times
and calculate the mean of the results. Compare the obtained means to the expected means.
If not, contact Mindray customer service department or your local distributor for assistance.
Other operations
Printing new calibration factors
Press [PRINT] to print out the current calibration factors.
Exiting the “Manual” screen
Press [MENU] to exit to the system menu.