6 - 6
Syringe Near
Empty
Low The preset Time Near
End is reached.
The alarm is cleared when
the infusion is completed.
End the infusion or replace
the syringe.
Battery
Depleted
High The battery is depleted. Connect the pump to the
external power source.
VTBI
Complete
High The preset VTBI is
completed.
Press to reset the
alarm.
Continue therapy or select
new therapy.
KVO Finish High The KVO infusion is
running for thirty
minutes.
Press to reset the
alarm.
Continue therapy or select
new therapy.
Relay Invalid High The pump is
disconnected from
the Dock.
In the relay state, the
upstream pumps
have completed the
infusions yet the
downstream pumps
are not ready for
infusion.
Check the connection
between the pump and the
Dock.
Check that the downstream
pumps are properly
configured for infusion.
System Error High The pump system faults,
such as storage error,
hardware fault, etc.
Stop using the pump, and
contact your service personnel.
KVO Running Low The infusion is
completed and the
pump continues
infusion at the KVO rate.
The alarm is cleared after the
KVO infusion reaches 30
minutes.
Press to pause the KVO
infusion.
Complete the infusion or
prepare for a new therapy.
Battery in Use Low The external power
source has been
disconnected and the
pump runs on battery
power.
Press to reset the
alarm.
Connect the pump to the
external power source.
Alarm Priority Causes Solutions