7 - 1
7 Menu Options
7.1 Main Menu Options
The main menu includes the following options:
7.2 General Option
The General Option menu comprises a list of options for configuring the infusion. To
access the General Option menu, follow this procedure:
1. Swipe the touchscreen from top down→ select Menu → select General Option.
2. Select the desired option.
Menu Item Details
General Option See 7.2 General Option.
System Options See 7.3 System Options.
Patient Management See 10 Managing Patient.
Discharge Patient See 10 Managing Patient.
Day's Prescription See 5.3 Performing Day’s Prescription.
User Maintenance See 11 Password Protected Settings.
Dock Setup See 5.4 Configuring Pumps in Batches through the Dock.
Menu Item Default Range Function
Current
Pressure
/ / Displays the current line pressure.
OcclusionPres
sure
450mmHg See A.7 Infusion
Specifications.
Set the occlusion alarm limit. The
pump gives the Occlusion alarm
when the occlusion pressure exceeds
the alarm limit.
KVO Rate 0.5ml/h See A.7 Infusion
Specifications.
Set the KVO rate. If KVO rate is set to
zero, the pump stops infusion when
VTBI is completed.