12 - 6
NOTE
If the battery is not conditioned for a prolonged time, its charge indication
may not be accurate and you may wrongly evaluate the remaining battery
runtime.
Do not use the pump for infusion during battery conditioning.
Do not interrupt battery conditioning.
12.5 Checking Version Information
To view the system software version, brand library version, drug library upgrade time,
and Wi-Fi version, follow this procedure:
1. Swipe the touchscreen from top down select Menu select User
Maintenance input the required password select
.
2. Select Version Information.
12.6 Checking the History Record
The History Record menu shows the history of pump activities, including the infusions,
alarms, calibrations, maintenance configurations, and other operations.
To access the History Record menu, follow this procedure:
1. Swipe the touchscreen from top down select Menu select System Options.
2. Select History Record.
NOTE
A total loss of power has no impact on the history records stored.
Alarms are saved as events and will remain if the equipment is powered
down. The time of equipment power down is also recorded as an event.
The pump stores up to 3500 events. When the capacity is reached, earlier
events will be overwritten by later ones.
12.7 Exporting the History Record
To export the history record, follow this procedure:
1. Connect the USB drive to the USB connector.
2. Swipe the touchscreen from top down select Menu select System Options.