C-1
C Alarm Information
This chapter presents the alarm information of the infusion pump. Prompt
information for operation guidance will not be presented in this chapter.
The table shows the appropriate countermeasures for each piece of information
related to alarm triggering. If the problem still exists after operating according to the
countermeasures, please contact the company.
Infusion cannula blocked
during infusion, and
occlusion pressure reaches
the threshold of preset
occlusion pressure threshold.
Infusion is stopped after the
alarm is triggered.
Press to cancel
alarm, check and
eliminate the source of the
alarm.
Size of single air detection
reaches to the preset value.
Infusion is stopped after the
alarm is triggered.
Press to cancel
alarm, check and
eliminate the air in line.
Size of accumulate air
bubble accumulated in 1
hour reaches to the preset
value. Infusion is stopped
after the alarm is triggered.
Press to cancel
alarm, check and
eliminate the air in line.
If drop sensor is installed
correctly and the switch of
[Drop sensor] is on, drop
sensor detects that the drop
rate deviates the preset
value. Infusion is stopped
after the alarm is triggered.
Press to cancel
alarm, refill the drop
chamber correctly or place
the drop chamber
vertically or avoid the
strong light.
1. Only powered with built-in
battery, battery is empty.
2. Built-in battery and
external DC power encounter
failure.
Connect to the AC power
source to cancel alarm.