3-4
WARNING
The infusion pump must be installed horizontally.
NOTE
Please ensure the stand is stable before installation.
When the pumps are overlapped, all fixing clamps except the fixing
clamps of the lowest pump should be removed, otherwise, the pumps
can't be used overlapped.
3.1.5 Mount the Handle Set
Align the rail on both sides of the handle set to the rail slot on the top of the lower
pump; slide from backward to frontward until the alignment is matched. When
disconnecting, toggle the button of the handle set outward and pull out the handle set
or pump.
Press the button of pole clamp, horizontally or vertically adjust pole clamp, turn the
handle, pump can be fixed to cross bar or vertical bar.
Square tube: Width 10mm, length 15-40mm
Round tube: External diameter ø15~ ø40