4-7
3. Please refer to 4.2.2.1 Infusion Pump with Conventional Infusion Sets to
reload the infusion set.
4.2.3.2 Infusion Pump with Silicone Specified Infusion Sets
Follow the steps below to change the silicone specified infusion set:
1. To prevent patient injury due to free flow, before changing the infusion set, please
shut down the Robert clip (or flow rate regulator). During infusion, press to
stop the pump.
2. Pull the door latch up to open the door, and push the door downward till the liquid
check clip of the infusion set completely falls off from door hook.
3. Press the liquid check clip operating lever of the infusion pump, take out the liquid
check clip and clips of the infusion set, and take out the loaded infusion set.
4. Please refer to 4.2.2.2 Infusion Pump with Silicone Specified Infusion Sets to
reload the infusion set.
4.2.3.3 Infusion Pump with PVC Specified Infusion Sets
Follow the steps below to change the PVC specified infusion set:
1. To prevent patient injury due to free flow, before changing the infusion set, please
shut down the Robert clip (or flow rate regulator). During infusion, press to
stop the pump.
2. Pull the door latch up to open the door, and push the door downward till the liquid
check clip of the infusion set completely falls off from door hook.
3. Press the liquid check clip operating lever of the infusion pump, take out the liquid
check clip of the infusion set, and take out the loaded infusion set.
4. Please refer to 4.2.2.3 Infusion Pump with PVC Specified Infusion Sets to
reload the infusion set.
4.2.4 Change the Infusion Bottle (Bag)
Follow the steps below to change the infusion bottle (bag):
1. To prevent patient injury due to free flow, before changing the infusion bottle (bag),
please shut down the Robert clip (or flow rate regulator). During infusion, press
to stop the pump.
2. Take out the loaded infusion bottle (bag), and reload it.
4.2.5 Select Infusion Set Brand
On the [Set selection] screen, press to select the infusion set brand, and
press for confirmation.