5-5
5.5 Loading Dose Mode
Loading Dose Mode infuses [VTBI] of loading dose at the rate of [Loading Rate],
then operates at [Main rate] until [VTBI] of the primary parameters is completed.
Primary parameters [VTBI] are [Total VTBI].
Mode Parameters Parameter Range
Loading
Dose
Mode
Loading dose
parameters
Volume, Rate and Time
The same with Rate
Mode
Primary
parameters
VTBI, Rate and Time
The same with Rate
Mode
NOTE:
1. Total VTBI should be greater than loading VTBI, and if set over the specified
limit, the parameters that exceed the system limit can not be set.
2. When any two parameters of the loading dose are null, the pump will run at the
[Rate] of primary parameters until the infusion set is empty or infuison is
completed.
5.6 Micro-infusion Mode
Clinically, the Micro-infusion Mode is mainly used for the infusion of children and
newborns at low rate.
Mode Parameters Parameter Range
Micro-infusion
Mode
Rate 0.1-100ml/h
VTBI 0.1-1000ml
Time 00:00:01-99:59:59 h:m:s