6-6
6.10.3 Infusion Pump with PVC Specified Infusion Sets
1. Select [Main Menu]→[User Maintenance]→Input User Maintenance
Password→[Infusion Set].
2. Select in [Infusion Set] according to actual needs.
Note: Please ensure that at least one “Infusion Set” to be selected.
Brand and model of PVC specified infusion sets:
1 Regular SK SK ZPQ
2 Regular SK SK 150mL
3 Light-sensitive SK SK JMB
6.11 Set the Sensitivity of Empty Bottle Alarm
1. Select [Main Menu]→[General Options]→[Empty Sensitivity].
2.
Select [Empty Sensitivity]: High and Low. [High] indicates high sensitivity of
[Empty] alarm detection, [Empty] alarm will be triggered in shorter time; [Low]
indicates low sensitivity of [Empty] alarm detection, [Empty] alarm will be triggered
in longer time. It can be selected according to actual needs
.
Note: [Empty Sensitivity] option is not visible when the switch of [Drop sensor] is
off.