C-3
Alarm
Information
Alarm
Level
Reason Countermeasure
[Relay Invalid] High
1. Upstream channel has not
start infusion or the infusion is
unfinished, the communication
with BeneFusion DS5 Infusion
Supervision System is
interrupted;
2. Or the upstream channel
has not start infusion or the
infusion is unfinished, the
communication between the
downstream channel and
BeneFusion DS5 Infusion
Supervision System is
interrupted;
3. Or the upstream channel
finishes infusion, the
downstream channel is not in
parameter setting interface, or
not set valid parameters for
starting infusion in parameter
setting interface.
Press or insert
infusion pump to
BeneFusion DS5 Infusion
Supervision System.
[System
Abnomal]
Medium
Power supply voltage
abnormal (low) and other
abnormal.
Alarm cannot be
cancelled. Please stop
operation and contact the
company.
[Standby Time
Expired]
Medium
Pump is in standby mode and
standby time is completed.
Press or to
cancel alarm.
[Reminder] Low
The infusion pump performs
no operation during the set
reminder time after the
infusion set is loaded.
1. Operate the pump
(except press ) to
cancel alarm.
2. Or press to
acknowledge the alarm
condition.
[Low Battery] Low
Only powered with built-in
battery, battery charge is
insufficient.
1. Connect to the AC
power source to cancel
alarm.
2. Or press to