Product Specifications
A-18
Microstream CO
2
Module
Standard Meet standard of EN864 and ISO9918
CO
2
Measurement range 0 to 99 mmHg
Precision*
0 to 38 mmHg:
39 to 99 mmHg:
±2 mmHg
±5% of the reading+0.08% of (the
reading-38)
* Accuracy applies for respiration rate up to 80 rpm. For respiration rate above 80 rpm, the
accuracy complies with EN 864/ISO 9918 (4 mmHg or ±12% of the reading, whichever is
greater) for EtCO
2
exceeding 18 mmHg. For respiration rate above 60 rpm, the above accuracy
can be achieved by using the CapnoLine H Set for Infant/Neonatal. In the presence of interfering
gases, the above accuracy is maintained to within 4% according to EN864 Section Eleven, Part
101.
Resolution 1 mmHg
Initialization time 30 s (typical)
Response time
2.9 s (typical)
(The response time is the sum of the rise time and the delay time
when using a FilterLine of standard length)
Rise time: <190ms (10% to 90%)
Delay time: 2.7s (typical)
awRR measurement range 0 to 150 rpm
awRR measurement accuracy
0 to 70 rpm:
70 to 120 rpm:
121 to 150 rpm:
±1 rpm
±2 rpm
±3 rpm
Apnea alarm time 10s, 15s, 20s, 25s, 30s, 35s, 40s
Mainstream CO
2
Module
Standard Meet standard of EN864 and ISO9918
CO
2
Measurement range 0 to 150 mmHg
Accuracy
0 to 40 mmHg:
41 to 70 mmHg:
71 to 100 mmHg:
101 to 150 mmHg:
±2 mmHg
±5% of the reading
±8% of the reading
±10% of the reading
Resolution
0 to 69 mmHg:
70 to 150 mmHg:
0.1 mmHg
0.25 mmHg
Response time <60ms
awRR measurement range 0 to 150 rpm
awRR measurement accuracy 1 rpm