6 Image Acquisition
Operator’s Manual 6 - 19
Scale
This function is used to adjust the speed range of color flow, which is adjusted through PRF in the
system. The real-time PRF value is displayed in the image parameter area.
To provide a much clearer color flow image.
Use low PRF to observe low-velocity flows, and use high PRF to observe high-velocity flows.
Aliasing may occur if low velocity scale is used and high velocities are encountered.
Low velocities may not be identified when a high velocity scale is used.
iTouch
+
To optimize image parameters as per the current tissue characteristics for a better image effect.
Auto Calculation
This function is used to trace the spectrum and calculate the PW/CW mode image parameters. The
results are displayed in the results window.
In real-time scanning, the results displayed are derived from the calculation of the latest cardiac
cycle.
In the freeze and cine status, the results displayed are calculated from the current selected area.
Auto Calc Param: To set the calculation results to display.
Auto Calc Cycle: To set the heart cycle number for auto-calculation.
Auto Calc Loop: To select the next loop or the last loop.
Trace Area: To set the trace area of the Doppler wave in the spectrum map, applicable for auto
calculation, V Max and V Mean display.
Trace Smooth: To set the smooth level when tracing.
Trace Sensitivity: This function is used to set the sensitivity of tracing in the spectrum.
Invert
This function is used to set the display manner of spectrum.
NOTE:
It is available only for linear probes.
Speed
This function is used to set the scan speed of PW mode imaging.
Changing the speed makes it easier to identify the cardiac cycles and to detect more details.
T/F Res
Adjusts for a balance between time resolution and spatial resolution.
WF (Wall Filter)
To display the image accurately, it adjusts the cut-off used in the wall filter, and filters out the flow
noise which is produced by vessel wall vibration. The real-time value is displayed in the image
parameter area.
Flow signals may be missing.
Tint Map
This function provides an imaging process based on color difference rather than gray distinction.