G EMC Guidance and Manufacturer’s Declaration
Operator’s Manual G - 7
Table G-6 Test specifications and minimum distances
Recommended separation distances between portable and mobile RF communications
equipment and system
Consona N6 series is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of system can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and Consona N6 series as recommended below, according to the maximum output power of the
communications equipment. Portable and mobile radio communications equipment (e.g. two-way radio,
cellular/ cordless telephones and similar equipment) should be used no closer to any part of this system,
including cables, than determined according to the following method:
Test
frequency
(MHz)
Band(MHz) Service Modulation Maximum
power (W)
Distance
(m)
Immunity
test level
(V/m)
385 380 - 390 TETRA 400 Pulse
modulation
18Hz
1.8 0.3 27
450 430 -470 GMRS 460
FRS 460
FM ±5 kHz
deviation 1
kHz sine
20.328
710 704 - 787 LTE Band 13,17 Pulse
modulation
217 Hz
0.2 0.3 9
745
780
810 800 - 960 GSM 800/900,
tetra 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
20.328
870
930
1720 1700 -1990 GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE Band 1, 3, 4,
25, UMTS
Pulse
modulation
217 Hz
20.328
1845
1970
2450 2400 -2570 Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450, LTE
Band 7
Pulse
modulation
217 Hz
20.328
5240 5100 -5800 WLAN,
802.11 a/n
Pulse
modulation
217 Hz
0.2 0.3 9
5500
5785

Table of Contents