7 - 12 Operator’s Manual
7 3D/4D
Reset Curve
Render Mode
The rendering manners can be applied to both gray and inversion modes.
Function: to inverse the echo of the 3D image, so as to enhance observation for low-echo region,
applicable for vessel, cyst and etc.
When the function is turned on, the rendering mode parameters change into the corresponding
inverse parameters.
VR Refine Optimize the signal-noise ratio and the contrast of VR image.
MagiClean Reduce the noise and keep the boundary of the image more clearly.
Available only in Surface rendering mode.
Depth VR Superimpose the tint map basic on the VR image to improve the stereoscopic
sensation and the contrast of the image.
Hyaline • Function: Improve the transparency in iLive rendering mode to observe
the tissue, such as cyst.
• Adjust method: The merging ratio of two render modes between Hyaline
and iLive is 100%.
• Effect: Adjust the merging ratio of two render modes between Hyaline and
iLive
Thickness Adjust the rendering thickness of the section.
Parameter Description
Parameter Description
Reset Ori To reset the volume rotation, shifting and zooming of 3D
image to original status.
All To reset the parameters, rendering rotation, VOI and image
effect.
Curve To reset the curve to be the original beeline.
Surface Set Surface as 3D image rendering mode.
Applicable for surface imaging, such as fetus face/hand or foot.
NOTE:
You may have to adjust the threshold to obtain a clear body boundary.
Max Set Max as 3D image rendering mode, displays the maximum echo intensity in
the observation direction.
This is helpful for viewing bony structures.
Min Set Min as 3D image rendering mode, displays the minimum echo intensity in
the observation direction.
This is helpful for viewing vessels and hollow structures.
X Ray Set X-ray as 3D image rendering mode. Displays the average value of all gray
values in the ROI.
This is used for imaging tissues with different structure inside or tissues with
tumor.