17 Probes and Biopsy
Operator’s Manual 17 - 23
a. Screw out the nut on the guiding block to loose the guiding block and the needle-guided
bracket.
b. Take the guiding block away in the direction of the needle tail, and then separate the
needle from the probe and the residual parts of the needle-guided bracket.
c. Screw out the locking nut on the needle-guided bracket to loose the bracket with the
probe.
d. Hold the probe with one hand, and then separate the needle-guided bracket from the
probe.
NGB-006 Plastic/needle detachable needle-guided bracket
No. Name Description
1. Support of needle-
guided bracket
Used for installing the needle-guided bracket on the probe.
2. Angle block Used for determining the angle of the biopsy; there are three
specifications of blocks of angle.
3. Guiding block Used for installing biopsy needle; there are five specifications
of guiding blocks for different biopsy needles.
4. Groove and tab of the
needle-guided bracket
Respectively matches with the tab and groove of the probe.