17 Probes and Biopsy
Operator’s Manual 17 - 37
5. Remove the needle-guided bracket:
a. Screw the pinch nut to release the needle-guided bracket.
b. Separate the bracket and the probe.
NGB-025
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Open the clamp. Insert the front clamp to the front groove, and align front clamp with
front groove.
1 Pinch nut 2 Intra-cavity probe
3 Lower clamp 4 Location hole
5 Front slot 6 Front clamp
7 Location clamp 8 Upper clamp
1
2
34
5
678