17 Probes and Biopsy
Operator’s Manual 17 - 47
c. Arrange the probe sheath in order. Press the left and right clamps of the bracket to
crosswise clamp the probe. Check if the bracket is properly installed to ensure that the it
won’t come off.
d. Hold the probe and insert the needle through the horn-shaped opening at the end of the
bracket into needle guide.
2. Remove the needle-guided bracket:
a. Remove the needle from the needle guide.
b. Hold the probe in one hand, and hold the bracket with the other hand. Press the separating
button of the bracket to open the clamp.