Loading...
17 - 48 Operator’s Manual
17 Probes and Biopsy
c. Hold the clasp at the end of the bracket to separate the bracket from probe.
NGB-052
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Hold the probe in one hand, select the correct needle-guided bracket and hold it with the
other hand. Match the inside of the bracket with the outside of the probe respectively.
Mount the bracket onto the probe. Rotate the locking knob of the guided bracket at the
right and left sides to fix the bracket and the probe. Rotate the angle adjusting knob to
adjust the probe guide structure to a proper angle, and then tighten the angle adjusting
button.
1 Probe fixing structure 2 Separating knob
3 Needle guide structure 4 Angle adjusting knob
5 Clamp 6 Locking knob
7 Guiding block
1
2
3
4
6
5
7

Table of Contents