4 - 28 Operator’s Manual
4 Setup
a. Select [New Stage] in the Stage list.
b. Enter a phase name.
Select [Auto Select:], the system jumps to Select Mode after retrospective acquisition.
Select the required option from the Clip Capture drop-down list.
Select the number of loops to acquire (per view in the selected phase) in the Loops list (for
non-continuous stages).
Select the type: exercise or drug.
5. Click [OK] to save changes and quit.
Editing a Stress Echo Protocol
Perform the following procedure:
1. Click a user-defined protocol on the Select Protocol screen.
2. Edit the protocol as described in the create protocol.
Deleting a Stress Echo Protocol
Click an user-defined protocol on the Select Protocol screen, click [Delete Protocol].
4.6.2 Maintenance
4.7 DICOM/HL7
4.7.1 DICOM Local Preset
NOTE:
AE Title should be the same with the SCU AE Title preset in the server (PACS/RIS/HIS), for
example, if the AE Title of the server preset in the storage server is Storage, and the AE Title of
the accepted SCU is preset as Machine, then in the figure above, the AE Title of Local should
be Machine, and the AE Title of storage server should be Storage.
Item Description
Acquire Mode Set the type of ROI: manual ROI or full-screen.
Overlay Select the items to be labeled on each loop.
WMS Score
Type
Set the chamber segment division method.
QT-Time Table To customize the length of systolic duration acquired for a specific heart rate, it
will store the clip duration. You can add and remove entries in this table. You
can also load the factory defaults.
Heart rate Enter the heart rate.
Syst. duration Enter the systolic duration.
Load Factory To reset the QT time table.
Update Enter a heart rate and the referring systolic duration and then click [Update].
Delete Select the required heart rate and systolic duration pair from the QT – Time
Table and then click [Delete].